System  
Measurement techniques: single and dual wavelength digital holographic microscopy in reflection and vertical coherence scanning
Image types: intensity and quantitative phase contrast image (DHM mode), optical topography (vertical scanning)
Light sources: two monochromatic laser sources
Sample stage: manual or automated XYZ stages up to 150 mm × 150 mm × 12.5 mm travel range
Camera: 1392 x 1040 pixel, 8 bits
Available objectives: standard, high NA, long working distance, water/oil immersion microscope objectives
Objective mounting: 4-position turret
Computer: Dell workstation with latest Intel ® processor, optimized and configured for DHM, with 19” SXGA monitor
Software: Lyncée Tec proprietary Koala classic software based on a C++ and .NET
Optional working mode: stroboscopic mode
   
Performance  
Measurement mode:
  Single
wavelength
Dual wavelength Vertical scanning
Accuracy 1 :
0.1 nm   25 nm (0.1 nm) 4 0.5 µm  
Vertical resolution 2 :
0.2 nm   50 nm (0.2 nm) 4 1.0 µm  
Repeatability 3 :
< 0.01 nm   < 0.25 nm (< 0.01 nm) 4 < 0.05 µm  
Vertical calibration: determined by the wavelength, no mechanical movement calibration
Vertical measuring range in single wavelength: up to depth of field for smooth samples, up to 340 nm for sharp edge samples
Vertical measuring range in dual wavelength: up to depth of field for smooth samples, up to 10 µm for sharp edge samples
Vertical measuring range in vertical scanning: up to 10 mm, z-stage dependent
Lateral resolution: objective dependent, down to 300 nm with oil immersion objectives (1.4 NA)
Field of view: objective dependent up to 4.4 mm
Working distance: objective dependent from 0.3 mm to 18 mm
Digital focusing range: up to 50× depth of field (objective dependent)
Grabbing time (1 hologram): down to 1 µs in a single image grab
Spatial sampling: 1024 x 1024 pixels (hologram)
Acquisition rate: 15 fps (1024 × 1024 pixels) (optional up to 300 fps)
Single wavelength reconstruction rate: 15 fps (512 × 512 pixels), 4 fps (1024 × 1024 pixels)
Dual wavelength acquisition time: 1.5 s
Vertical scanning acquisition time: scanning speed: 6 µm/s, reconstruction time: 6 s
Min. sample reflectivity: less than 1%
Sample illumination: down to 1 µW/cm2
   
Power requirements  
Input voltage: 85-260 VAC - 50/60 Hz
Power requirements (w/o computer): max 480 W
   
Dimensions & weight  
Microscope:
R1101 R1102
Dimensions (L × W × H):
600 × 600 × 800 mm 900 × 900 × 850 mm
Weight:
45 kg 80 kg
Maximum sample size (L × W):
8 × 415 mm 530 × 870 mm

Specifications are subject to change without notice

1   As demonstrated by taking the temporal standard deviation on 1 pixel over 30 measurements*.
2   Defined as twice the accuracy.
3   As demonstrated by taking the one sigma Rq value of 30 repeatability measurements* on SiC reference mirror.
4   When combined to single wavelength measurements.
*  1 measurement is the average of 10 acquisitions.